Stairs.  Entry group.  Materials.  Doors.  Locks.  Design

Stairs. Entry group. Materials. Doors. Locks. Design

» Dzmo sterilization equipment. Sterilization boxes (medical bixes), round KF and KSKF Replacement filter for KF and KSKF boxes

Dzmo sterilization equipment. Sterilization boxes (medical bixes), round KF and KSKF Replacement filter for KF and KSKF boxes

Round sterilization box with filters type KF (Russia) and KSKF (Belarus)

Box with Filter (type: KF 3, KF 6, KF 9, KF 12, KF 18), And Round Sterilization Box with Filter(type: KSKF 3, KSKF 6, KSKF 9, KSKF 12, KSKF 18) or medical beaks for sterilization- absolutely similar in purpose and application, differ slightly in design and manufacturer - intended for placing in it items and materials for medical purposes such as: dressings, heat-resistant syringes, surgical linen, surgical material and other medical items for the purpose of sterilizing them in steam sterilizers and delivery to the place of use, as well as sterile storage of all contents for 20 days.
Sterilization bins (KF sterilization round box with filters) - made of corrosion-resistant, non-magnetic polished stainless steel (ISO 9001), recommended for use in medical equipment, equipped with replaceable filters made of cotton fabric (from filter diagonal article 2074/100 GOST 332- 91.)
Composition of the round sterilization box with filter (KF):
- Frame
- pen
- lid
- lock
- filter (2 pcs.)
- clamp
- latch
- earring
- retainer

Filters for sterilization boxes make it possible to sterilize medical objects and materials in steam sterilizers under pressure (0.2 ± 0.02) MPa, at a temperature of 132 ± 2 °C for 20 minutes or under pressure (0.11 ± 0. 02) MPa, at a temperature of 120 ºС for 45 minutes, while the filters are suitable for microbiological filtration of the air during subsequent storage. To replace used filters, the company produces sets of spare filters for sterilization boxes of all sizes. The set consists of two filters. One package contains 5 sets of filters for boxes of the same size. Sets of spare filters can also be used to replace filters in round sterilization boxes of the KF type. Since 2004, filters of the same standard size in KSKF and KF boxes are completely interchangeable!
Sterilization box KF and sterilization box KSKF - during operation it is prohibited to use without filters!
Bix medical KF has a warranty period of 36 months
Package: The KF box and the KSKF box are sold only assembled; inside the box there are instructions for use (manufacturer’s passport) and 2 filters. The sterilization box with assembled filters is packed in individual packaging containers (branded cardboard box).
Peculiarities: The sterilization box KF (medical box with filter) produced by OJSC "DZMO" (Doschatinsky Medical Equipment Plant) does not include any rubber seals in its design, since it is used special technology rolling sheet metal, and also has 3 on the base of the body metal legs which provide a guaranteed gap for the passage of steam when installing boxes in the sterilizer on top of each other - this is essential technical advantage over other box manufacturers.

Name KF-3 KF-6 KF-9 KF-12 KF-18
Diameter in mm. 190 (±5) 250 (±5) 290 (±5) 340 (±5) 390 (±5)
Height (with legs) in mm 140 (±10) 160 (±10) 160 (±10) 160 (±10) 190 (±10)
Conditional volume in dm³ 3 6 9 12 18
Product weight in kg. 0,6 1,0 1,2 1,6 2,0
Dimensions with packaging in mm:
width
length
height

179
179
149

235
235
165

285
285
165

335
335
165

380
380
200

Price in rubles 749,00 1110,00 1415,00 1615,00 1918,00

By purchasing sterilization boxes KF and KSKF note for the presence of: - the original manufacturer's stamp on the Certificate of Conformity - the original company factory mark on the product. Round sterilization boxes KSKF produced by OJSC "Grodno Trade Mechanical Engineering Plant", Belarus
differ from other manufacturers by the presence of a quick-release filter clamp, convenient latches - in the lid and at the bottom of the sterilization box. All products have a Russian Certificate of Conformity, as well as Registration Certificates from the Ministry of Health of Russia, Kazakhstan, Uzbekistan, and the Kyrgyz Republic.

Warranty period: 36 months

Sterilization box KSKF dimensions:

Specifications KSKF-3 KSKF-6 KSKF-9 KSKF-12 KSKF-18
Nominal volume (dm³) 3 6 9 12 18
Diameter in mm, no more 175 245 275 325 390
Height in mm, no more 170 170 190 190 190
Weight in kg, no more 0,75 1,15 1,45 1,9 2,45
Filter diameter (mm) 110 135 155 175 175
Number of locks 1 2 2 2 2
Price, in rub.

Replacement filter for KF and KSKF boxes

Filter for sterilization box KF and KSKF - made of thick cotton fabric specially designed for this purpose - filter diagonal article 2074/100 GOST 332-91. Fabric filter diagonal
It is produced by weaving equilateral four-shaft twill. The filters are suitable for reusable use and can be easily disposed of by burning; they are guaranteed to withstand at least 60 sterilization cycles - they are a consumable replacement material, so additional filters of all sizes can be purchased separately. The set of filters for the KF and KSKF boxes consists of 2 pieces. Filters for KSKF sterilization boxes are installed in special places on the bottom and lid of the box. The KF filter is placed under a clamp and snapped into place with a clamp (KSKF) or under a screw clamp (KF). Sets of spare filters can also be used to replace filters in round sterilization boxes CF. Since 2004, filters of the same standard size in KSKF and KF boxes are interchangeable.

If the universal filter is larger seat on the sterilization box - its edges (excess) are cut off with ordinary scissors, as well as the hole for the fastening (the filter is folded in half and the tip is cut off).

Also purchased with this product:

Round sterilization boxes with filters

KF-3/6/9/12/18

Sterilization boxes KF-3/6/9/12/18 are intended for placing medical items and materials in it for the purpose of sterilization in steam sterilizers, further storage and delivery to the place of use.

Principle of operation boxes is based on the penetrating ability of steam through the filters inside the boxes and the ability of the filters to microbiologically filter the air environment.

Basic bridles and details

The boxes consist of housings with hinged lid, which is tightly closed with locking locks. The hinged joint allows the lid to be opened as fully as possible.

The cover is equipped with a pen to carry the box.

On the base of the case there are legs

Boxes are completed two replaceable filters from filter diagonal art. 2074 according to GOST 332, which are mounted in the lid and on the bottom, with inside boxes using clamps.

The period of preservation of sterility in the box after sterilization is up to 20 days according to MU-287-113.

The service life of filters with one sterilization per day is 1 month.

Replacement filters from other manufacturers cannot guarantee sterility.

Characteristics:

* overall diameter includes hinge and lock

** overall height is indicated from the top of the cover for the bottom, taking into account the length of the legs

Advantages:


2. At the base of the box body there are legs that provide a guaranteed gap for the passage of steam when installing boxes with a steam sterilizer on top of each other.

Sterilization boxes KSPF-9/KSPF-18 are intended for placement of medical products made of corrosion-resistant metals, glass, rubber, latex, certain types of plastics, textile products, as well as dressings and suture materials, for the purpose of sterilization in steam sterilizers, delivery to the place of use in a sterile state and storage after sterilization within a regulated period. Used in hospitals and other medical institutions.
Principle of operation
boxes is based on the penetrating ability of steam through the filters inside the box and the ability of the filters to microbiologically filter the air environment.
Main components and parts
are made from of stainless steel.
The body is provided pens for carrying the box.
On the base of the case there are legs, providing a guaranteed gap for the passage of steam when installing boxes on top of each other in the sterilizer.
Lid
— removable, equipped with a sealing silicone gasket, securely fixed to the body using two locks.
The design of the box allows for installation fillings to detect the fact that the box has been opened.
Boxes are completed replaceable filters from filter diagonal art. 2074 according to GOST 332, which are fastened in the lid and on the bottom, on the inside of the box, using clamps.
Sterility retention period
in a box after sterilization up to 20 days according to MU-287-113.
Filter life
for 1 month with 1 sterilization per day.

Characteristics:

Advantages:

1. The boxes are made of high quality non-magnetic stainless steel.
2. Rectangular shape boxes allows you to install them as tightly as possible in the steam sterilizer.
3. The presence of a sealing silicone gasket on the lid ensures a tight fit of the lid to the body of the box.
4. A handle for carrying the box is provided silicone tube what protects medical staff from the possibility of burns.
5. At the base of the box body there are legs that provide a guaranteed gap for the passage of steam when installing boxes with a steam sterilizer on top of each other.

Disinfection boilers

Designed for disinfection of instruments by boiling in distilled water. The boiler is used in medical institutions .

Principle of operation The boiler is based on electric heating and boiling water in a container with a tool placed in it.

Frame The boiler, tray and lid are made of stainless steel. The main parts and components of the boiler are made of corrosion-resistant materials.

Control block consists of an input box with an electrical cord, fuses, a power-on indicator and the heating device. The power cord is equipped with a two-pole plug with a grounding contact.

The body is equipped with a stainless steel tray with perforation to accommodate disinfection objects. The tray is removed from the body using plastic grips.

When plugged in, the indicator lights up and water heating begins.

The boiler automatically turns off when the water boils over or there is no water.

The boiler is supplied in corrugated cardboard packaging.

Characteristics:

Advantages:

1. The boiler is made of high-quality non-magnetic stainless steel, 1mm thick.
2. The boiler is equipped with a perforated tray.
3. The tray is removed from the boiler using plastic grips, which protects medical personnel from the possibility of burns.

STERILIZATION BOXES- containers for placing dressings, surgical linen, surgical instruments, gloves and other items during their sterilization, delivery to the place of use and subsequent short-term storage (boxes). S. to. allow to create in the treating-prof. institutions reserve sterile materials for urgent surgical interventions and promote the centralized use of sterilization equipment (see Scientific organization of labor).

In the second half of the 19th century. in Moscow, G.V. Mendel proposed the use of zinc boxes for placing dressings in them during sterilization. Later, E. Braatz proposed wire baskets (Braatz boxes) for instruments during autoclaving, and A. Diihrssen proposed cylindrical tin boxes with holes around the perimeter, in the bottom and in the lid. K. Schimmelbusch used a tetrahedral box with a lattice bottom (Schimmelbusch box) to sterilize materials.

Currently, round and rectangular filters are used, with and without a filter. Their main specifications are given in the table.

Round sterilizers without a filter (bixes) are intended for use in sterilizers (see), which have a predominantly cylindrical sterilization chamber. They consist of a body, a cover and a belt moving along the body (Fig. 1). There are holes in the body and in the belt, which are aligned during the sterilization period (before placement in the sterilizer). When unloading the S. from the sterilizer, the belt is shifted, closing it. holes so that the sterile instruments and materials located in the SK do not come into contact with the air of the room. The sterility of the material is maintained for up to 3 days. The belt is equipped with a tension lock, which prevents its spontaneous movement.

For more long-term preservation sterility of materials (up to 20 days) are used. round type KSKF (Fig. 2) and rectangular type KSPF (Fig. 3). They are equipped with biol. a filter made of textile materials (diagonal filter, non-woven cloth, bleaching calico, calico, madapolam, etc.), which are attached to the inside of the body and lid.

It is recommended to place a gasket on the bottom of the container before placing materials in it ( cotton fabric, gauze or other material), which will absorb the moisture that accumulates after sterilization. It is better to wrap individual items (tools, gloves) in a double layer of gauze before placing them in the S.C. and place them loosely (on the edge) in the S.C. Linen and dressings should be rolled into flat bags and placed in the S.C. without compaction. also on the edge so that the palm fits freely between them. When closing the S. k., the lid is locked. The lid should not compress the contents of the sterilizer. When sterilizing, to control the temperature in the sterilizer, they use samples with substances whose melting point is known (sulfur melts at 115°, benzoin at 122°, etc.). These substances in the form of powder in bottles or test tubes are placed in the sterilizer among the material being sterilized, and at the end of sterilization they check whether the powder has melted or not. Sometimes indicators are added to the powder - safranin, brilliant green, etc., which color them at the moment of melting in the corresponding color. A tag is attached to the lock of the closed lid, which indicates the date of sterilization, the name of the employee who performed the installation, a list of sterilization objects located in the sterilization center, and in the case of a centralized sterilization facility, the department.

Table. MAIN TECHNICAL CHARACTERISTICS OF DOMESTIC STERILIZATION BOXES

Box code

External dimensions, mm

Weight (mass), kg

With filter

Without a filter

Diameter (for round ones), length and width (for rectangular ones)

With filter

Without a filter

Rectangular

are issued

are issued

* The diameters of sterilization boxes produced before 1981 are given in parentheses.